当前位置:首页 > 产品展示 > > 耐压测试仪,安规测试仪 > 7742安规综合分析仪
详情介绍
7742安规综合分析仪,全新六合一功能包含 (ACW,DCW,IR,GB,LLT,RUN TEST)
全新六合一功能包含 (ACW,DCW,IR,GB,LLT,RUN TEST)
超大型320×240 Graphic LCD 窗口
可于同一画面下完整的显示所有的设定参数或量测数值
测试功能确认设定(Verification)
自动提示讯息功能(Prompt)
快速放电线路 (:104990)
可设定校正日期-时间的预警功能
内建50组程序记忆、每一组30个测试步骤,可于一次设定连续测试
全范围电压输出稳压率在1%以内
键盘锁定功能
内含防高压触电线路Smart GFI (:169000)
简单易懂的人机界面
缓升上限(Ramp High)设定(:100859)
充电下限(Charge Low)设定 (:106128)
内建PLC Remote可程序遥控界面
可外接标准GPIB/RS-232界面作控制
备有小电流接地阻抗检查功能(Continuity)
量测速度快,准确度高
可选购SPACE-9170安规自动化测试软件,经由个人计算机PC作控制
| |||||||
AC WITHSTAND VOLTAGE | |||||||
Range | Resolution | Accuracy | |||||
Output Voltage, ACV | 0-5000 | 1 | ± (2% of setting + 5V) | ||||
Output Frequency | 50Hz/60Hz ±0.1%, User Selection | ||||||
Output Waveform | Sine Wave ,THD.<2% (Resistive Load), Crest Factor = 1.3 - 1.5 | ||||||
Output Regulation | ± (1% of output + 5V), From no load to full load | ||||||
SETTINGS | |||||||
HI and LO-Limit (Total), current, mA | 0.000-9.999 (0=OFF) | 0.001 | ± (2% of setting + 2 counts) | ||||
10.00-40.00 | 0.01 | ||||||
HI and LO-Limit (Real), current, mA | 0.000-9.999 | 0.001 | ± (3% of setting + 50µA) | ||||
10.00-40.00 | 0.01 | ||||||
Ramp-Up Timer, second | 0.1 - 999.9 | 0.1 | ± (0.1% + 0.05 sec) | ||||
Ramp-Down Timer, second | 0.0-999.9 | ||||||
Dwell Timer, second | 0, 0.4 - 999.9 (0=continuous) | ||||||
Ground Continuity | Current: DC 0.1 A ± 0.01A, fixed | ||||||
Max. Ground Resistance: 1 Ω ± 0.1Ω, fixed | |||||||
Arc Detection | 1 - 9 ranges (9 is the most sensitivity) | ||||||
DC WITHSTAND VOLTAGE | |||||||
Output Voltage, DCV | 0 -5000 | 1 | ± (2% of setting + 5V) | ||||
DC Output Ripple | <4 %(5KV/20mA at Resistive Load) | ||||||
SETTINGS | |||||||
HI and LO-Limit, current, µA | 0.0 - 999.9 | 0.1 | ± (2% of setting + 2 counts) | ||||
1000 - 20000 | 1 | ||||||
Ramp-Up Timer, second | 0.4 - 999.9 | 0.1 | ± (0.1% + 0.05 sec) | ||||
Ramp-Down Timer, second | 0.0, 1.0-999.9 | ||||||
Dwell Timer, second | 0, 0.3 - 999.9 (0=continuous) | ||||||
Ramp-HI, current | >20 mA peak maximum, ON/OFF User Selection | ||||||
Charge-LO, current | 0.0 - 350.0 µA DC or Auto Set | ||||||
Discharge Time | ≦ 200 msec | ||||||
Maximum Capacitive Load DC Mode | 1µF < 1KV 0.08µF < 4KV | ||||||
0.75µF < 2KV 0.04µF < 5KV | |||||||
0.5µF < 3KV | |||||||
Ground Continuity | Current: DC 0.1 A ± 0.01A, fixed | ||||||
Max. Ground Resistance: 1 Ω ± 0.1Ω, fixed | |||||||
Arc Detection | 1 - 9 ranges (9 is the most sensitivity) | ||||||
INSULATION RESISTANCE | |||||||
Output Voltage, DCV | 50 - 1000 | 1 | ± (2% of setting + 2 counts) | ||||
Charging Current | Maximum >20mApeak | ||||||
SETTINGS | |||||||
HI and LO-Limit, resistance, MΩ | 0.05 - 99.99 (HI – Limit: 0 = OFF) | 0.01 | Same as Resistance MEASUREMENT Accuracy | ||||
100.0-999.9 | 0.1 | ||||||
1000-50000 | 1 | ± (0.1% + 0.05 sec) | |||||
Ramp-Up Timer, second | 0.1 – 999.9 | 0.1 | |||||
Ramp-Down Timer, second | 0=OFF, 1.0-999.9 | ||||||
Delay Timer, second | 0, 1.0 - 999.9 (0=continuous) | ||||||
Charge-LO, current, µA | 0.000 - 3.500 or Auto Set | ||||||
GROUND BOND | |||||||
Output AC Current, A(7742 ) | 1.00 – 40.00 | 0.01 | ± (2 % of setting + 2 counts) | ||||
Output Voltage, Vac | 3.00 – 8.00 | 0.01 | ± (2 % of setting + 3 counts) | ||||
Output Frequency, Hz | 50Hz/60Hz ±0.1%, User Selection | ||||||
Output Regulation | ± (1% of output + 0.02A), Within maximum load limits, and over input voltage range. | ||||||
Maximum Loading | 1.00–10.00A/0–600mΩ, 10.01–30.00A/0–200mΩ, 30.01–40.00A/0–150mΩ | ||||||
SETTINGS | |||||||
Lead Resistance Offset, mΩ | 0-200 | 1 | 1.00-2.99A,± (3 % of setting + 3 counts ) 3.00 - 30A,± (2 % of setting + 2 counts ) 3.00 - 40A,± (2 % of setting + 2 counts ) | ||||
HI and LO-Limit Resistance, mΩ | 0 – 150 (30.01-40.00A) | 1 | |||||
0 – 200 (10.01-30.00A) | |||||||
0 – 600 (1-10.00A) | |||||||
Dwell Timer, second | 0, 0.5 - 999.9 (0 = continuous) | 0.1 | ± (0.1% + 0.05 sec) | ||||
CONTINUITY TEST | |||||||
Output Current, DC | 0.1A ± 0.01A, fixed | ||||||
SETTINGS | |||||||
Lead Resistance Offset, Ω | 0.00-2.00 | 0.01 | ± (3% of setting+2 counts) | ||||
HI and LO-Limit, Resistance, Ω | 0.00 – 10.00 (0=OFF) | ||||||
Dwell Timer, second | 0,0.3-999.9 | 0.1 | ± (0.1% + 0.05 sec) | ||||
Opt.734 (0 - 10kΩ) | |||||||
Output Current | DC 0.1A ± 0.01A,Max. | ||||||
SETTINGS | |||||||
Max-Lmt, Ω | 0.00 – 9.99 | 0.01 | ± (1% of setting+5 counts) | ||||
10.0 – 99.9 | 0.1 | ± (1% of setting+2 counts) | |||||
100 – 999 | 1 | ||||||
1000 – 9999 | 1 | ||||||
Min-Lmt, Ω | 0.00 – 9.99 | 0.01 | ± (1% of setting+5 counts) | ||||
10.0 – 99.9 | 0.1 | ± (1% of setting+2 counts) | |||||
100 – 999 | 1 | ||||||
1000 – 9999 | 1 | ||||||
Dwell Timer, second | 0.0, 0.3 – 999.9 (0 = continuous) | 0.1 | ± (0.1% + 0.05 sec) | ||||
Resistance Offset, Ω | 0.00 – 2.00 | 0.01 | ± (1 % of reading + 5 counts) | ||||
MEASUREMENT | |||||||
Range | Resolution | Accuracy | |||||
Voltage, KV(AC/DC) | 0.00-5.00 | 0.01 | ± (1.5 % of reading +1 count) | ||||
Voltage, Vdc (IR only) | 0-1000 | 1 | ± (1.5% of reading + 2 counts) | ||||
AC Current (Total), mA | 0.000-3.500 | 0.001 | ± (2% of reading + 2 counts) | ||||
3.00-40.00 | 0.01 | ||||||
AC Current (Real), mA | 0.000-9.999 | 0.001 | ± (3% of reading + 50?A) | ||||
10.00-40.00 | 0.01 | All Ranges PF > 0.1 , V > 250VAC | |||||
DC Current, µA | 0.0 - 350.0 | 0.1 | ± (2% of reading + 2 counts) | ||||
DC Current, mA | 0.300 – 3.500 | 0.001 | |||||
3.00-20.00 | 0.01 | ||||||
AC Current, A (G.B for 7742 only) | 0.00-40.00 | 0.01 | ± (3 % of reading + 3 counts) | ||||
Resistance, MΩ (IR) | 50-499V / 500-1000V | | 50 – 499V0.05–999.9, ±(7% of reading+2 counts) | ||||
0.050-1.999 / 0.050-9.999 | 0.001 | 500 – 1000V | |||||
2.00-19.99 / 0.050-9.999 | 0.01 | 0.10–999.9, ±(2% of reading+2 counts) | |||||
20.0-199.9 / 100.0-999.9 | 0.1 | 1000–9999, ±(5% of reading+2 counts) | |||||
200-50000 / 1000-50000 | 1 | 10000–50000, ±(15% of reading+2 counts) | |||||
Resistance, mΩ (GB) | 0-600 | 1 | 1.00 – 2.99 A, ± (3 % of reading + 3 counts) | ||||
3.00 – 40.00 A, ± (2 % of reading + 2 counts) | |||||||
Resistance, Ω (Continuity) | 0.00-10.00 | 0.01 | ± (3 % of reading + 2 counts) | ||||
Resistance, Ω (Continuity) Opt.734 | 0.00 – 9.99 | 0.01 | ± (1% of reading+5 counts) | ||||
10.0 – 99.9 | 0.1 | ± (1% of reading+2 counts) | |||||
100 – 999 | 1 | ||||||
1000 – 9999 | 1 | ||||||
GENERAL | |||||||
Input Voltage AC | 115/230Vac±15%, 50/60Hz ± 5%, Fuse 10A/250V Slow-Blow for 7742 Model | ||||||
PLC Remote Control | Input : Test, Reset, Interlock, Recall File 1 through 10 | ||||||
Output:- Pass, Fail, Test-in-Process | |||||||
Memory | 50 memories, 30 steps/memory | ||||||
Graphic Display | 320 x 240 Monographic LCD, 9 ranges contrast setting | ||||||
Safety | Built-in Smart GFI circuit,GFI trip current 450µA max, HV shut down speed: <1mS | ||||||
Interface | Standard RS232,Optional GPIB (IEEE-488.2),Multi-function Interface Card,Ethernet Card. | ||||||
Security | Programmable password lockout capability to avoid unauthorized access to test set-up program. | ||||||
Alarm Volume Setting | Range: 0-9 ;0=OFF, 1 is softest volume, 9 is loudest volume. | ||||||
Calibration | Software and adjustments are made through the front panel. Automatic Calibration alert function to signal operator when re-calibration is due. | ||||||
Environment | 0-40℃, 20-80%RH | ||||||
Dimensions/Net Weight | 430mm(W)×133mm(H)×500 mm(D)/30Kg for MODEL 7742 | ||||||
STANDARD ACCESSORIES | |||||||
Power Cord(10A) | ×1 | ||||||
Fuses | ×2 (Including a spare contained in the fuse holder) | ||||||
Interlock Disable Key(1505) | ×1 | ||||||
Hipot Test Lead ,1.5m(1101) | ×1 | ||||||
Ground Bond Test Lead 40A ,1.6m(1137) | ×1 | ||||||
Ground Bond Return Lead 40A ,1.6m(1138) | ×1 | ||||||
RS232 Link Cable 1.5m(1130) | ×1 | ||||||
MATRIX SCANNER MODULE (Opt.736 for 7742) | |||||||
High Voltage Rating | 5KVAC/5KVDC | ||||||
High Current Rating | 40 A | ||||||
Number of HV Channel | 8 | ||||||
Number of HA Channel | 8 | ||||||
RUN TEST MODULE (Opt.738,Opt.739 for 7742) | |||||||
MEASUREMENT | |||||||
Range | Resolution | Accuracy | |||||
Voltage, Vac | 00.-277.0 | 0.1 | ± (1.5% of reading +2counts), 30.0-277.0V | ||||
Current, Aac | 0.00-15.00 | 0.01 | ± (2.0% of reading + 2counts) | ||||
Power, Watts | 0-4200 | 1 | ± (5% of reading + 3 counts) | ||||
Power, Factor | 0.000-1.000 | 0.001 | ± (8% of reading + 2 counts) | ||||
Leakage Current, mA | 0.00-10.00 | 0.01 | ± (2% of reading + 2 counts) | ||||
MD Circuit | Leakage Current measuring resistance = 2KΩ±1% | ||||||
Timer, second | 0.0-999.9 | 0.1 | ± (0.1% of reading + 0.05 sec) | ||||
SETTINGS | |||||||
HI and LO Limit AC Voltage, V | 0.0-277.0 | 0.1 | ± (1.5% of setting + 0.2 V), 30.0-277V | ||||
HI and LO Limit AC Current, A | 0.00-15.00 | 0.01 | ± (2.0% of setting + 2 counts) | ||||
HI and LO Limit AC Power, W | 0 - 4200 | 1 | ± (5.0% of setting + 3 counts) | ||||
HI and LO Limit Power Factor | 0.000-1.000 | 0.001 | ± (8% of setting + 2 counts) | ||||
HI and LO Limit Leakage Current | 0.00-10.00, Hi-Limit 0=OFF | 0.01 | ± (2% of setting + 2 counts) | ||||
Delay Time, second | 0.2-999.9 | 0.1 | ± (0.1% + 0.05 sec) | ||||
Dwell Time, second | 0, 0.1-999.9 (0=continuous) | ||||||
Cold Resistance | 0~10KΩ | ||||||
DUT POWER | |||||||
AC Voltage | 0 - 277.0V, Single phase unblance, 0 - 15.0A maximum | ||||||
Power Rating | 4200 W maximum | ||||||
DUT Protection | Short circuit current 23A, <3 sec, Inrush Current 68A Response time 1m sec | ||||||
STANDARD ACCESSORIES | |||||||
Power Cord(15A) | ×1 | ||||||
LLT Receptacle Adaptor Box 20A ,1.6m(1905) | ×1 | ||||||
LINE LEAKAGE TEST MODULE (Opt.739 for 7742) | |||||||
CURRENT MEASUREMENT | |||||||
Frequency Range | DC-1MHz | ||||||
Leakage Current Range (RMS) | Resolution | Accuracy (DC-100KHz) | Accuracy ( >100KHz to 1 MHz) | ||||
0.0µA - 999.9µA | 0.1µA | ±(1.5% of reading + 3counts) | ±(5% of reading + 5µA) | ||||
1000µA - 6000µA | 1µA | ±(1.5% of reading + 3counts) | |||||
Leakage Current Range (Peak), Opt.739 Only | Resolution | Accuracy (DC-100KHz) | Accuracy ( >100KHz to 1 MHz) | ||||
0.0µA - 999.9µA | 0.1µA | ±(10% of reading + 2µA) | ±(10% of reading + 2µA) | ||||
1000µA - 6000µA | 1µA | ||||||
LINE VOLTAGE MEASUREMENT | |||||||
Range | 0.0-277.0 VAC | ||||||
Resolution | 0.1 V | ||||||
Accuracy | ± (1.5% of reading +0.2V), 30.0-277.0V | ||||||
DUT | |||||||
DUT Input Power Rating | 0-277V,AC@ 15Aac max | ||||||
DUT Protection | Short circuit current 23A, <3 sec, Inrush Current 68A Response time 1m sec | ||||||
SETTINGS | |||||||
High/Low Limit, µA | Range 0.0 - 6000µA(0=OFF) , Resolution 0.1µA | ||||||
Delay Time, second | Range 0, 1.0-999.9 (0=continuous), Resolution 0.1second | ||||||
Compliance Standards | A UL544 Non Patient, UL484, IEC60598, UL1363 | ||||||
B UL544 Patient Care | |||||||
C UL2601-1, IEC60601-1, EN60601-1 | |||||||
D UL1563 | |||||||
E UL60950, IEC60950, IEC61010-1, IEC60335-1, IEC60990 Fig4 U2 | |||||||
F IEC60990 Fig5 U3 | |||||||
G IEC60990 Fig3 U1 | |||||||
MD A - G components | Accuracy: Resistance ± 1% Capacitor ± 5% | ||||||
MD Voltage Limit | Maximum 30V peak or 30VDC | ||||||
STANDARD ACCESSORIES | |||||||
Hipot Return Lead ,1.8m(1102) | ×2 |